SA Instruments has recently released a new PC-SAM software version (11.02) for Models 1030, 1025T and 1025 and recommends users to update to this latest version. The update includes a corrected ERT thermistor temperature reading for ERT Model 1030.
For ERT 1030 Modules with firmware version 38, previous software versions display systematically too high (Version < 10.03) or too low (10.3 ≤ Version < 11.00) temperature readings. The offset is in the range of 0.7°C at 37°C.
For temperature measures obtained with different PC-SAM versions, the corrected temperature (T1c) can be calculated from the measured temperature (T1):
Please note: We recommend to complete running studies before performing the software update.
PC-SAM 11.02 is available from the vendor website for download on https://i4sa.com/ (external link).
With a range of positioning systems to perfectly match instrument and individual needs, as well as animal cradles for basic and specialized applications, set-ups have never been faster or more straightforward.