Probes

Micro2.5 – MicWB40

The Micro2.5 / MicWB40 probe is the perfect tool for imaging of a wide range of small objects. The MicWB40 rf probe is equipped with exchangeable coils supporting sample diameters from 1 to 30 mm.

The Micro2.5 / MicWB40 probe is the perfect tool for imaging of a wide range of small objects. The MicWB40 rf probe is equipped with exchangeable coils supporting sample diameters from 1 to 30 mm. If required it can be equipped with the animal monitoring system to supervise physiological parameters of the animals (ECG, respiration, body temperature) and provide connections and masks for breathing gas anesthesia. Different type of animal beds and positioning systems are supported for the individual coil types. An optional VT set allows to control the temperature of the sample between -20°C and +60°C.

The Micro2.5 gradient system offers a gradient strength of up to 1.5 T/m and is perfectly suited for most applications from small ex vivo samples to in vivo experiments with mice.

Features

  • Wide Bore Probe, compatible to SWB magnets
  • Gradient strength of up to 1.5 T/m per axis, water cooled
  • Animal support (Triggering Unit, Anesthesia)
  • Exchangeable Rf coils, named ‘MicWB40’ series
    - 2.5 mm Solenoid
    - 5-30 mm Single Tuned Coils, in part circular polarized
    - 5-30 mm Double Tuned Coils
  • Surface Coils
  • Optional Variable Temperature Control (compatible to BSVT, BCU,…)
  • Special sample holders for glass tubes or animal beds are available. 

Specifications

MicWB40  
Bore     WB / SWB
1H frequency range 300 – 800 MHz
rf channels Two independent
Requirements MR Microscopy Accessory
Gradient     Micro2.5    
Directions     xyz    
Gradient strength per direction 1.5 T/m
ID/OD 40/72 mm
Rise Time (5-95%), 0-60A,120V < 100 µs
Cooling     Water    
Maximum current tested 60 A
Removable     Yes    
Exchangable rf coil  
Detection     Single tuned, double tuned, or quadrature
Typical Nuclei 1H, 2H, 13C, 19F, 31P,....
Standard sample temperature range (5 mm) -20 – 60°C
Sample diameter 1 – 30 mm

Accessories

  • 2.5 mm solenoid coil. Due to the B1 direction, the samples must be placed horizontally. 
  • 5 – 30 mm volume coil (in 5 mm steps). The sample needs to be placed manually into the rf coil. Single tuned coils up to 15 mm are linear coils, while 20-30 mm coils are circular polarized coils (needs quad combiner or two receive/transmit channels in the console). Double tuned coils (5-30 mm) are always linear. Comes with sample holder in corresponding diameter.
  • Surface coils (10, 15, 20 mm) can be tuned to a single frequency where the object needs to be placed directly on the coil and is typical used for in-vivo experiments. The animal handing is included with the coil. 
  • Animal Handling Accessory
  • Animal Monitoring Unit 